Which Iv Medications Need A Filter To Administer

Introduction

Air or particles introduced into the patient circulation via venous access can have catastrophic consequences for patients who have communications between their arterial and venous circulations. Venous emboli that cross from the right side (venous circulation) to the left side of the heart into the arterial circulation bypass filtering through pulmonary capillary diffusion into alveoli and therefore have greater proximity to the coronary and cerebral circulations. Hence, there is a greater need for filtering emboli for patients with mixed venous and arterial circulations. Using 0.2micron in-line filters can help prevent air, particle, and microbe introduction into the patient circulation via intravenous lines.

Aim

  • To identify cardiac patients who are at an increased risk of systemic emboli from air or particles entrained into venous access
  • To provide information on the management of in-line filters to minimise the risk of embolisation in these patients
  • To demonstrate the correct position of in-line filters in venous access lines

Management

Cardiac patients who require filters

  • In-line filters must be placed in all direct atrial lines.
  • In-line filters must be placed in all arterial and all central and peripheral intravenous lines in the following circumstances:
    • Pre-operative or unrepaired patients with a congenital heart defect with communications between their venous and arterial circulations.
    • Pre and post-operative patients with single ventricle lesions.
Refer to more articles:  Which Process Primarily Transfers Heat By Moving Air

Filters available

A 0.2micron filter filters particulate debris, microbial contaminants and their endotoxins and entrained air. Once the membrane of the filter is wet during priming, air is no longer able to enter the patient.

Two sizes are used at the Royal Children’s Hospital.

1. Pall Nanodyne®NEO Filter (Figure 1)

Priming volume of 0.4mL

Maximum flow rate of 110mL/hr

Should be used for neonates and infants

2. Pall Nanodyne ® ELD Filter (Figure 2)

Priming volume of 2.6mL

Maximum flow rate of 1260-1380 mL/hr

Both filters need to be changed every 96 hours (4 days) to assure microbe and endotoxin retention.

Filters administering parenteral nutrition must be changed every 24 hours.

Fig 1 Pall Posidyne NEO filter

Figure 1. Pall Nanodyne ®NEO filter

Fig 2 Pall Posidyne ELD filter

Figure 2. Pall Nanodyne ® ELD filter

Pall Nanodyne® NEO 0.2micron filters (Figure 1) can be used for Epoprostenolol (Prostacyclin) infusions. This filter is to be changed every 24 hours.

Placement and use of filters

Identify in advance the patients likely to require an in-line filter before their admission to the unit or arrival from theatre.

For postoperative cardiac patients requiring inotrope infusions, filters should be placed onto the intravenous lines by the anaesthetist at the commencement of the infusion in theatre. When this has not occurred, the infusion should be changed to a venous line with a filter as soon as the patient’s condition allows.

Multiple infusions that are compatible for administration via one lumen and do not require bolus (e.g., inotropes line) can be infused through a shared filter where the filter should be attached to a needless adaptor as close as possible to the patient. See example in Figure 3.

Refer to more articles:  Which Of The Following Statements About Quorum Sensing Is False

Fig 3 neo filter

Figure 3. One in-line filter is required near the patient hub while multiple inotropes can be infused distally via three-way taps

Volume and maintenance CVAD lumen

In the event a fluid bolus must be administered, an in-line filter would impede quick and easy administration. Despite the risk of emboli, an in-line filter cannot be used in the venous line used for filling. Volume should be administered via a three-way tap or ‘traffic light’ system and may added to the CVAD maintenance line if required (See Figure 4). Other drug ports and the maintenance lines that share a lumen with the filling line should have their own individual filters.

Fig 4 cvad lumen setup

Figure 4. Set up of maintenance CVAD lumen (with volume line if required)

Invasive monitoring or direct atrial lines in ICU

Link to invasive haemodynamic cardiac monitoring guideline HERE

Place the filter between the syringe and transducer, so that it does not interfere with the pressure reading (see Figure 5). Filters may result in a slightly dampened waveform trace. Turning the three-way tap to temporarily isolate the transducer from the infusion may assist a more detailed waveform analysis.

Fig 5 neo filter pressure monitoring

Figure 5. Set up of pressure monitoring lines in PICU

Medications that cannot be filtered

Most medications and fluids commonly administered to cardiac patients can be filtered through the 0.2 micron filter.

Medications with limited evidence that cannot be filtered through a 0.2micron filter include:

  • Amphotericin
  • Asparaginase
  • Blood products
  • Cyclosporin
  • Lipid
  • Octreotide
  • Propofol

Substances that cannot be filtered should ideally be infused by themselves. In situations in which this is not possible, the unfiltered infusion should be attached as close to the patient as possible. All other compatible infusions should be infused through a filter.

Refer to more articles:  Which Is More Dangerous Jamaica Or Dominican Republic

In the event a patient requires a CT with contrast, ensure contrast is administered through a venous line without a filter as the filter will delay the delivery of contrast.

Troubleshooting

If you notice increasing pressure on your infusion pump(s) or occlusion alarms, this may be due to a blocked filter. A blocked filter does not mean the filter has failed. Rather it’s more likely the filter has succeeded in stopping debris from entering the patient’s circulation. A VHIMS listing the fluids/drugs infusing through the filter and their infusing rates should be completed. This will add to the body of information about infusion (in)compatibilities.

Companion Documents

  • Cardiac set up for PICU (RCH only access)
  • Cardiac post-operative patient reception in PICU (RCH only access)
  • CVAD management guideline – Insertion, Management and Removal of Central Venous Access Devices
  • Invasive haemodynamic cardiac monitoring (RCH only access)

Evidence Table

The evidence table for this guideline can be viewed here.

Please remember to read the disclaimer. The revision of this nursing guideline was coordinated by Sathika Tilakaratne, RN, PICU, approved by the Nursing Clinical Effectiveness Committee. Published January 2023.

Related Posts

Which Ankle For Ankle Bracelet

Which Ankle For Ankle Bracelet

An ankle bracelet, or anklet, can be worn with a casual, formal, or any other kind of outfit. Slip an ankle bracelet on whenever you want to…

Which Is Not A Function Of Epithelial Tissue

What is the epithelium? The epithelium is a type of body tissue that forms the covering on all internal and external surfaces of your body, lines body…

Which Animal Has The Best Memory

Memory is the brain’s ability to encode and recall the needed information. In other words, memory – the capacity, which enables the remembrance of events, thought, feeling,…

Which Of The Following Genotypes Is Homozygous

What is homozygous? In genetics, the definition of homozygous is when you inherit the same DNA sequence for a specific gene from each of your biological parents….

Which Lines Are Parallel Check All That Apply

Which Lines Are Parallel Check All That Apply

Parallel lines are the lines that do not intersect or meet each other at any point in a plane. They are always parallel and are at equidistant…

Which Of The Following Is True Of Bloodborne Pathogens

Under OSHA’s Bloodborne Pathogens Standard requirements, all employees are to be regularly trained on the risks of occupational exposure to infectious materials. Workers should have a deep…